Amorphous selenium (a-Se) flat-panel digital mammography detectors are being investigated for tomosynthesis, which poses tremendously challenges on the detector temporal and low dose performance. Our previous investigation has demonstrated that a-Se detectors provide adequate temporal performance (lag and ghosting) for tomosynthesis, however its detective quantum efficiency (DQE) at 1 mR (1/10 of average exposure in screening mammography) was only ~1/3 of the value at 10 mR due to electronic noise. Before engineering methods can be developed to overcome this problem, optimization of imaging parameters, such as x-ray spectrum and anti-scatter grid, can greatly improve the detector performance at the low dose used in tomosynthesis. The purpose of this paper is to determine the optimal x-ray spectrum and whether an anti-scatter grid is beneficial in tomosynthesis. The SNR of a 200 μm microcalcification within the breast was calculated as a function of x-ray spectra. Two target materials (Mo and W) were used. The density and thickness of the breast were varied. The scatter to primary ratio behind the breast with and without grid was calculated. The detector performance of a state-of-the-art a-Se digital mammography detector with 85 micron pixel size was incorporated in the calculation of SNR. The total breast dose was kept constant at 1.6 mCy. Our results showed that for tomosynthesis with 11 acquisition views, the optimal kVp is at least 2-3 kVp higher than the optimum for screening mammography. In the extreme case of an 8 cm dense breast, the optimal spectrum was 39 kVp (W/Rh), which was 9 kVp higher than the optimal kVp when detector noise is negligible. W/Rh was found to be the optimal target filter combination for all breast thicknesses (2-8 cm). Our results also showed that grid has no clear advantage even for breast thickness of 8 cm.
机译:目前正在研究非晶硒(a-Se)平板数字乳房X线照像仪探测器的断层合成,这对探测器的时间和低剂量性能提出了极大的挑战。我们之前的研究表明,a-Se检测器可为断层合成提供足够的时间性能(滞后和重影),但是其在1 mR(乳腺X线摄影术中平均曝光量的1/10)下的检测量子效率(DQE)仅为1/3由于电子噪声,该值在10 mR时的最大值。在开发能够克服该问题的工程方法之前,优化成像参数(例如X射线光谱和防散射网格)可以极大地提高断层合成中使用的低剂量检测器的性能。本文的目的是确定最佳的X射线光谱以及反散射网格是否对层析合成有益。乳房内200μm微钙化的SNR计算为X射线光谱的函数。使用了两种目标材料(Mo和W)。乳房的密度和厚度各不相同。计算了有无网格的情况下,乳腺背后的散射与主要比例。在SNR的计算中结合了具有85微米像素大小的先进a-Se数字乳房X线摄片机的检测器性能。乳房的总剂量保持恒定在1.6 mCy。我们的结果表明,对于具有11个采集视图的断层合成,最佳kVp比乳腺钼靶筛查的最佳值至少高2-3 kVp。在8厘米密集乳房的极端情况下,最佳频谱为39 kVp(W / Rh),比检测器噪声可忽略时的最佳kVp高9 kVp。发现W / Rh是所有乳房厚度(2-8厘米)的最佳目标过滤器组合。我们的结果还显示,即使胸围8厘米,网格也没有明显的优势。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号