Loss of Power Output and Laser Fibre Degradation during 120 Watt Lithium-Triborate HPS Laser Vaporisation of the Prostate

机译:前列腺的120瓦三硼酸锂HPS激光汽化过程中的功率输出损失和激光纤维降解

获取原文
获取原文并翻译 | 示例

摘要

It has recently been shown that laser fibre deterioration leads to a significant decrease of power output during 80 W potassium titanyl phosphate (KTP) laser vaporisation (LV) of the prostate. This decrease results in inefficient vaporisation especially towards the end of the procedure. For the new 120 W lithium-triborate (LBO) High Performance System (HPS) laser not only higher power but also changes in beam characteristics and improved fibre quality have been advertised. However, high laser power has been identified as a risk factor for laser fibre degradation. Between July and September 2008 25 laser fibres were investigated during routine 120 W LBO-LV in 20 consecutive patients. Laser beam power was measured at baseline and after the application of every 25 kJ during the LV procedure. Postoperatively, the surgeon subjectively rated the performance of the respective fibre on a scale from 1 to 4 (1 indicating perfect and 4 insufficient performance). Additionally, microscopic examination of the fibre tip was performed. Median energy applied was 212 kJ. Changes of power output were similar for all fibres. Typically, a steep decrease of power output within the first 50 kJ was followed by a continuous mild decrease until the end of the procedure. After the application of 50 kJ the median power output was 63% (58-73% interquartile range) of the baseline value. The median power output at the end of the 275 kJ-lifespan of the fibres was 42% (40-47%). The median surgeons' rating of the overall performance of the laser fibres was 2 and the median estimated final decrease of power output 60%. Some degree of degradation at the emission window was microscopically detectable in all cases after the procedure. However, even after the application of 275 kJ, these structural changes were only moderate. Minor degradation of the laser fibre was associated with a significant decrease of power output during 120 W LBO-LV. However, following an early, steep decrease, power output remained relatively constant on a medium level for the rest of the fibre's lifespan. The subjective rating by the surgeons is in accordance with these findings. Improved properties of the LBO laser and enhanced fibre quality resulted in an only moderate decrease of power output which allowed for a consistently effective performance.
机译:最近显示,在前列腺的80 W钛氧基磷酸钾(KTP)激光汽化(LV)期间,激光纤维的劣化会导致功率输出显着下降。这种减少导致低效的汽化,特别是在程序快要结束时。对于新型的120 W三硼酸锂(LBO)高性能系统(HPS)激光器,不仅宣传了更高的功率,而且还改变了光束特性并改善了光纤质量。但是,高激光功率已被确定为激光纤维降解的危险因素。在2008年7月至2008年9月之间,在20例连续患者中常规行120 W LBO-LV期间研究了25根激光光纤。在基线过程中以及在LV程序中每施加25 kJ后测量激光束功率。术后,外科医生主观评价各个纤维的性能,等级为1至4(1表示完美,4表示性能不足)。另外,对纤维尖端进行了显微镜检查。施加的中值能量为212 kJ。所有光纤的功率输出变化均相似。通常,在开始的50 kJ内功率输出急剧下降,然后连续缓慢下降直到过程结束。施加50 kJ后,中值功率输出为基准值的63%(四分位间距的58-73%)。光纤在275 kJ寿命结束时的平均功率输出为42%(40-47%)。外科医生对激光纤维总体性能的评价的中位数为2,估计的功率输出最终下降的中位数为60%。在该程序之后的所有情况下,在显微镜下都可检测到发射窗口处的某种程度的降解。但是,即使在施加275 kJ的压力后,这些结构变化也只是中等程度。激光纤维的轻微降解与120 W LBO-LV期间的功率输出显着降低有关。但是,在早期急剧下降之后,在光纤的剩余寿命中,功率输出在中等水平上保持相对恒定。外科医生的主观评价与这些发现一致。 LBO激光器性能的提高和光纤质量的提高导致功率输出仅适度下降,从而实现了始终如一的有效性能。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号