首页> 外文会议>Photonic Therapeutics and Diagnostics II; Progress in Biomedical Optics and Imaging; vol.7, no.1 >A Comparison of Single Use and Reusable Small Core Sized Holmium:YAG Laser Fibers: Is There a Difference in the Risk of Fiber Failure?
【24h】

A Comparison of Single Use and Reusable Small Core Sized Holmium:YAG Laser Fibers: Is There a Difference in the Risk of Fiber Failure?

机译:一次性使用和可重复使用的小型Core:YAG激光光纤的比较:光纤失效的风险是否有所不同?

获取原文
获取原文并翻译 | 示例

摘要

This study tests the hypothesis that repeat steam sterilization will result in an increase in the rate of holmium:YAG laser fiber failure during bench testing in a series of commercially available single use and reusable small core sized (200-272 μm) holmium:YAG laser fibers. Single use and reusable small core-sized holmium laser fibers were tested. Single use fibers included the Dornier Lightguide Super 200, Dornier DUR Laser Fiber Single Use, Optical Integrity Scopesafe 272, IQinc. LLF200TG-D and LLF273TG-D, Boston Scientific AccuFlex 200 and AccuFlex 273, and Lumenis Slimline EZ 200. Reusable fibers included the Dornier DUR Laser Fiber Reusable, Lumenis Slimline 200, Sharplan 200, Laser Peripherals RBLF-200, IQinc. LFT273NT, and Convergent Optiview SMH1020F. A Lumenis VersaPulse 100 watt and a PowerSuite 20 watt holmium:YAG lasers were used. Fibers were bent to 180 degrees at a diameter of 1.5 cm. The laser was fired at 1.5J, 10Hz for 30 minutes or until fiber fracture. Reusable fibers were sterilized for a total of twenty cycles using the manufacturers' specifications and retested after every five completed cycles. No fiber fractured with bending alone. Two of three Dornier Super 200 single use fibers fractured repeatedly within several laser pulses. The Laser Peripherals RBLF-200 fiber fractured during initial but not repeat testing. One of three reusable Dornier DUR Laser Fibers failed during testing after the twentieth steam sterilization cycle. Five of five Boston Scientific Accuflex 200 failed at the SMA connector and did not couple well with the Lumenis Powersuite laser. The test hypothesis was not supported.
机译:这项研究检验了以下假设:在一系列市售的一次性使用和可重复使用的小尺寸core(YAG)激光(200-272μm)中,反复进行蒸汽灭菌将导致Y:YAG激光光纤故障在台式测试期间增加的失败率。纤维。测试了一次性使用和可重复使用的小型纤芯大小的激光光纤。一次性光纤包括Dornier光导Super 200,Dorrier DUR激光光纤一次性,光学完整性Scopesafe 272,IQinc。 LLF200TG-D和LLF273TG-D,Boston Scientific AccuFlex 200和AccuFlex 273以及Lumenis Slimline EZ200。可重复使用的光纤包括可重复使用的Dornier DUR激光光纤,Lumenis Slimline 200,Sharplan 200,激光外围设备RBLF-200,IQinc。 LFT273NT和Convergent Optiview SMH1020F。使用了Lumenis VersaPulse 100瓦和PowerSuite 20瓦:YAG激光器。将纤维以1.5cm的直径弯曲至180度。激光器以1.5J,10Hz的频率发射30分钟或直到光纤断裂。使用制造商的规格将可重复使用的纤维灭菌总共20个循环,并在每5个完整循环后进行重新测试。单独弯曲不会使纤维断裂。多尼尔Super 200一次性光纤中的三根中有两根在几个激光脉冲中反复断裂。激光外围设备RBLF-200光纤在初始测试过程中断裂,但没有重复测试。在第二十次蒸汽灭菌循环后的测试过程中,三支可重复使用的Dornier DUR激光光纤中的一根失败了。波士顿科学Accuflex 200中有五分之二的SMA连接器出现故障,并且与Lumenis Powersuite激光器耦合不好。不支持检验假设。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号