Comparison of Holmium:YAG and Thulium Fiber Lasers on the Risk of Laser Fiber Fracture

机译:钬:YAG和纤维激光激光激光纤维骨折风险的比较

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Objectives: To compare the risk of laser fiber fracture between Ho:YAG laser and Thulium Fiber Laser (TFL) with different laser fiber diameters, laser settings, and fiber bending radii. METHODS: Lengths of 200, 272, and 365 μm single use fibers were used with a 30 W Ho:YAG laser and a 50 W Super Pulsed TFL. Laser fibers of 150 µm length were also tested with the TFL only. Five different increasingly smaller bend radii were tested: 1, 0.9, 0.75, 0.6, and 0.45 cm. A total of 13 different laser settings were tested for the Ho:YAG laser: six fragmentation settings with a short pulse duration, and seven dusting settings with a long pulse duration. A total of 33 different laser settings were tested for the TFL. Three laser settings were common two both lasers: 0.5 J × 12 Hz, 0.8 J × 8 Hz, 2 J × 3 Hz. The laser was activated for 5 min or until fiber fracture. Each measurement was performed ten times. Results: While fiber failures occurred with all fiber diameters with Ho:YAG laser, none were reported with TFL. Identified risk factors of fiber fracture with the Ho:YAG laser were short pulse and high energy for the 365 µm fibers (p = 0.041), but not for the 200 and 272 µm fibers (p = 1 and p = 0.43, respectively). High frequency was not a risk factor of fiber fracture. Fiber diameter also seemed to be a risk factor of fracture. The 200 µm fibers broke more frequently than the 272 and 365 µm ones (p = 0.039). There was a trend for a higher number of fractures with the 365 µm fibers compared to the 272 µm ones, these occurring at a larger bend radius, but this difference was not significant. Conclusion: TFL appears to be a safer laser regarding the risk of fiber fracture than Ho:YAG when used with fibers in a deflected position.
机译:目的:比较HO:YAG激光和纤维激光器(TFL)之间激光纤维骨折的风险,具有不同的激光纤维直径,激光设置和纤维弯曲半径。方法:200,272和365μm的长度使用纤维,与30W HO:YAG激光器和50W超级脉冲TFL一起使用。还使用TFL测试150μm长度的激光纤维。测试五种不同越来越小的弯曲半径:1,0.9,0.75,0.6和0.45cm。对HO:YAG激光器进行测试总共13个不同的激光设置:脉冲持续时间短,脉冲持续时间短,脉冲持续时间长为7个碎片设置。为TFL测试了总共33种不同的激光设置。三个激光设置是常见的两个激光器:0.5J×12 Hz,0.8J×8 Hz,2 J×3 Hz。激光被活化5分钟或直至纤维骨折。每次测量都进行了十次。结果:虽然具有HO:YAG激光的所有纤维直径发生的纤维失效,但TFL报告无。鉴定HO的纤维骨折的危险因素:YAG激光器为365μm纤维的短脉冲和高能量(P = 0.041),但不适用于200和272μm纤维(P = 1和P = 0.43)。高频不是纤维骨折的危险因素。纤维直径似乎也是骨折的危险因素。 200μm纤维比272和365μm更频繁地破裂(p = 0.039)。与265μm纤维相比,与272μm的纤维相比,这些趋势较多,这些趋势在较大的弯曲半径上发生,但这种差异并不重要。结论:TFL似乎是关于纤维骨折风险的更安全的激光器,而纤维在偏转位置中的纤维使用时。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号