Optical alignment using a CGH and an autostigmatic microscope

机译:使用CGH和自动散光显微镜进行光学对准

获取原文
获取原文并翻译 | 示例

摘要

We show how custom computer generated holograms (CGH) are used along with an autostigmatic microscope (ASM) to align both optical and mechanical components relative to the CGH. The patterns in the CGHs define points and lines in space when interrogated with the focus of the ASM. Once the ASM is aligned to the CGH, an optical or mechanical component such as a lens, a well-polished ball or a cylinder can be aligned to the ASM in 3 or 4 degrees of freedom and thus to the CGH. In this case we show how a CGH is used to make a fixture for cementing a doublet lens without the need for a rotary table or a precision vertical stage.
机译:我们展示了如何使用自定义计算机生成的全息图(CGH)以及自动像散显微镜(ASM)来相对于CGH对准光学和机械组件。当与ASM的焦点进行询问时,CGH中的模式定义了空间中的点和线。一旦将ASM对准CGH,就可以将光学或机械组件(如透镜,抛光的球或圆柱体)与ASM对准3或4个自由度,从而对准CGH。在这种情况下,我们展示了如何使用CGH制作用于固结双合透镜的固定装置,而无需旋转工作台或精密的垂直工作台。