首页> 外文会议>Optical diagnostics and sensing XVII: toward point-of-care diagnostics >In vivo preclinical verification of a multimodal diffuse reflectance and correlation spectroscopy system for sensing tissue perfusion
【24h】

In vivo preclinical verification of a multimodal diffuse reflectance and correlation spectroscopy system for sensing tissue perfusion

机译:用于检测组织灌注的多峰漫反射和相关光谱系统的体内临床前验证

获取原文
获取原文并翻译 | 示例

摘要

In reconstructive surgery, impeded blood flow in microvascular free flaps due to a compromise in arterial or venous patency secondary to blood clots or vessel spasms can rapidly result in flap failures. Thus, the ability to detect changes in microvascular free flaps is critical. In this paper, we report progress on in vivo pre-clinical testing of a compact, multimodal, fiber-based diffuse correlation and reflectance spectroscopy system designed to quantitatively monitor tissue perfusion in a porcine model's surgically-grafted free flap. We also describe the device's sensitivity to incremental blood flow changes and discuss the prospects for continuous perfusion monitoring in future clinical translational studies.
机译:在重建手术中,由于继发于血块或血管痉挛的动脉或静脉通畅性受损,微血管游离皮瓣血流受阻可迅速导致皮瓣衰竭。因此,检测微血管游离皮瓣变化的能力至关重要。在本文中,我们报告了紧凑型,多峰,基于纤维的弥散相关和反射光谱系统的体内临床前测试的进展,该系统旨在定量监测猪模型手术移植的游离皮瓣中的组织灌注。我们还将描述该设备对增量血流变化的敏感性,并讨论在未来的临床转化研究中进行连续灌注监测的前景。