Design study of a high-resolution breast-dedicated PET system built from cadmium zinc telluride detectors

机译:由碲化镉锌探测器构建的高分辨率乳房专用PET系统的设计研究

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

We are developing a dual-panel cadmium zinc telluride (CZT) detector positron emission tomography (PET) system dedicated to breast imaging. The proposed system consists of two 4 cm thick 12 x 15 cm2 area CZT panels with adjustable separation, which can be put in close-proximity to breasts and/or axillary nodes and results in high photon sensitivity. In this project, the performance of the proposed CZT dual panel PET system was investigated using Monte-Carlo simulation. Expected system performance is predicted due to CZT''s superior energy resolution, the high photon sensitivity, and the uniform spatial resolution throughout field-of-view (FOV). The dependency of system reconstructed resolution on depth-of-interaction (DOI) resolution was also studied.
机译:我们正在开发专用于乳房成像的双面板碲化镉锌(CZT)检测器正电子发射断层扫描(PET)系统。拟议的系统由两块4 cm厚的12 x 15 cm 2 区域CZT面板组成,这些面板具有可调的间距,可以紧密贴近乳房和/或腋窝结节,并具有很高的光子敏感性。在该项目中,使用蒙特卡洛模拟研究了拟议的CZT双面板PET系统的性能。由于CZT出色的能量分辨率,高光子灵敏度和整个视场(FOV)一致的空间分辨率,因此可以预测预期的系统性能。还研究了系统重构分辨率对交互深度(DOI)分辨率的依赖性。