Design study of a high-resolution breast-dedicated PET system built from cadmium zinc telluride detectors

机译:由镉碲化镉探测器内置的高分辨率乳房专用宠物系统的设计研究

获取原文

摘要

We are developing a dual-panel cadmium zinc telluride (CZT) detector positron emission tomography (PET) system dedicated to breast imaging. The proposed system consists of two 4 cm thick 12×15 cm{sup}2 area CZT panels with adjustable separation, which can be put in close-proximity to breasts and/or axillary nodes and results in high photon sensitivity. In this project, the performance of the proposed CZT dual panel PET system was investigated using Monte-Carlo simulation. Expected system performance is predicted due to CZT's superior energy resolution, the high photon sensitivity, and the uniform spatial resolution throughout field-of-view (FOV). The dependency of system reconstructed resolution on depth-of-interaction (DOI) resolution was also studied.
机译:我们正在开发一种专用于乳房成像的双面板镉碲化铈(CZT)探测器正电子发射断层扫描(PET)系统。所提出的系统由两个4厘米厚的12×15cm {sup} 2区域Czt面板组成,具有可调节的分离,可以对乳房和/或腋窝节点进行近距离,并导致高光子敏感性。在该项目中,使用Monte-Carlo仿真研究了所提出的CZT双面板PET系统的性能。预期的系统性能是由于CZT的卓越能量分辨率,高光子敏感性和整个视野(FOV)的均匀空间分辨率(FOV)而预测的系统性能。还研究了系统重建分辨率对相互作用深度(DOI)分辨率的依赖性。