摘要

We report a novel electrophoretic separation method that is realizable in a 1 mm single microfluidic channel, powered by a standard 9V battery and can be used to complete a migration shift immunoassay in less than 30s. In our assay sample is continuously injected electrophoretically into a polyacrylamide sieving matrix with analyte fronts monitored to detect differentially migrating species. By removing an injection channel this technique eliminates injection dispersion and shrinks the device footprint. Resulting in high efficiency separations in short microfluidic channels with a power consumption of just 2.7 μW.
机译:我们报告了一种新型的电泳分离方法,该方法可在1 mm单微流体通道中实现,由标准9V电池供电,可用于在不到30s的时间内完成迁移转移免疫测定。在我们的测定中,样品通过电泳连续注入聚丙烯酰胺筛分基质中,并监测分析物前沿以检测差异迁移的物质。通过去除注入通道,该技术消除了注入分散并缩小了器件占地面积。在短的微流体通道中实现高效分离,功耗仅为2.7μW。