首页> 外文会议>ASME International Conference on Nanochannels, Microchannels, and Minichannels >DETERMINING THE COMPRESSIVE BEHAVIOR OF AN ULTRA-THIN ELASTOVISCOPLASTIC MEMBRANE USED IN SEALING A MICROCHANNEL HEMODIALYSER
【24h】

DETERMINING THE COMPRESSIVE BEHAVIOR OF AN ULTRA-THIN ELASTOVISCOPLASTIC MEMBRANE USED IN SEALING A MICROCHANNEL HEMODIALYSER

机译:确定用于密封微通道血液透析器的超薄弹性囊泡膜的压缩行为

获取原文

摘要

In past efforts, the conditions were determined for the hermetic sealing of a microchannel hemodialyser produced by pressing a hot-embossed polycarbonate microchannel lamina containing sealing bosses, against a 25 μm thick elastoviscoplastic hemodialysis membrane. In this paper, a procedure is developed for quantifying the compressive mechanical behavior of the membrane during loading. An energy balance, of the work of clamping to the strain energy within the membrane, is used to derive a model for predicting the force required to seal the membrane. The force model is used to predict the force required to press a boss a certain distance into the membrane, providing validation of the procedure for measuring the compressive properties of ultra-thin elastoviscoplastic membranes.
机译:在过去的努力中,确定通过将含有密封凸台的热压花的聚碳酸酯微通道薄膜压制含有密封凸台的微通道血液氧化剂的气密密封,反对25μm厚的弹性粒子塑料血液透析膜。在本文中,开发了一种方法,用于在装载过程中量化膜的压缩力学行为。夹紧到膜内的应变能的工作的能量平衡用于导出用于预测密封膜所需的力的模型。力模型用于预测将凸台压入膜的一定距离所需的力,提供用于测量超薄弹塑性膜的压缩性能的过程的验证。