首页> 外文会议>IEEE Biomedical Circuits and Systems Conference >An active TX/RX NMR probe for real-time monitoring of MRI field imperfections
【24h】

An active TX/RX NMR probe for real-time monitoring of MRI field imperfections

机译:有源TX / RX NMR探针,用于MRI现场缺陷的实时监控

获取原文

摘要

In this paper, we present a PCB-based active miniaturized MR field probe for real-time monitoring of the magnetization's phase evolution during magnetic resonance (MR) experiments. The data obtained with the presented sensor can be used to correct gradient field imperfections which uncorrected result in significant distortions in the reconstructed MR images. The presented active field probe consists of a susceptibility matched solenoidal MR coil and a complete homodyne transceiver. Thanks to the local generation of the radio frequency signal required for the excitation of the spin ensemble and the downconversion of the recorded MR signal to low frequencies, the proposed architecture significantly reduces the crosstalk between the probe head and the MR imaging object compared to existing designs. MR measurements performed in an ultra high field 9.4 T full-body scanner prove the compatibility of the presented sensor with commercial MR imaging systems and demonstrate its excellent MR phase tracking performance.
机译:在本文中,我们介绍了一种基于PCB的有源小型化MR场探针,用于在磁共振(MR)实验期间对磁化相进化的实时监测。用所提出的传感器获得的数据可用于校正未校正的劣化导致重建的MR图像中的显着扭曲的缺陷。所提出的活性场探针包括敏感性匹配的螺线管MR线圈和完整的Homodyne收发器。由于局部生成旋转集合的激励所需的射频信号和记录的MR信号到低频的下变频,所提出的架构与现有设计相比,显着减少了探头和MR成像对象之间的串扰。在超高场9.4T全体扫描仪中进行的MR测量证明了所提出的传感器与商业MR成像系统的兼容性,并展示其优异的MR相跟踪性能。