首页> 外文会议>IEEE Biomedical Circuits and Systems Conference >An active TX/RX NMR probe for real-time monitoring of MRI field imperfections
【24h】

An active TX/RX NMR probe for real-time monitoring of MRI field imperfections

机译:有源TX / RX NMR探头,用于实时监测MRI现场缺陷

获取原文

摘要

In this paper, we present a PCB-based active miniaturized MR field probe for real-time monitoring of the magnetization's phase evolution during magnetic resonance (MR) experiments. The data obtained with the presented sensor can be used to correct gradient field imperfections which uncorrected result in significant distortions in the reconstructed MR images. The presented active field probe consists of a susceptibility matched solenoidal MR coil and a complete homodyne transceiver. Thanks to the local generation of the radio frequency signal required for the excitation of the spin ensemble and the downconversion of the recorded MR signal to low frequencies, the proposed architecture significantly reduces the crosstalk between the probe head and the MR imaging object compared to existing designs. MR measurements performed in an ultra high field 9.4 T full-body scanner prove the compatibility of the presented sensor with commercial MR imaging systems and demonstrate its excellent MR phase tracking performance.
机译:在本文中,我们提出了一种基于PCB的有源微型MR场探头,用于在磁共振(MR)实验期间实时监控磁化的相变。利用所提出的传感器获得的数据可以用于校正梯度场缺陷,这些未校正的结果导致重构的MR图像中明显的畸变。提出的有源场探头由磁化率匹配的螺线管MR线圈和完整的零差收发器组成。与现有设计相比,得益于本地产生的自旋系综激励和将记录的MR信号降频转换为低频所需的射频信号,所提出的体系结构大大减少了探头和MR成像对象之间的串扰。在超高视场9.4 T全身扫描仪中进行的MR测量证明了该传感器与商用MR成像系统的兼容性,并证明了其出色的MR相跟踪性能。