首页> 外文会议>IEEE Asian Solid-State Circuits Conference >A 20V-compliance implantable neural stimulator IC with closed-loop power control, active charge balancing, and electrode impedance check
【24h】

A 20V-compliance implantable neural stimulator IC with closed-loop power control, active charge balancing, and electrode impedance check

机译:具有闭环功率控制,主动充电平衡和电极阻抗检查的20V-Compliance Implantable神经刺激器IC

获取原文
获取外文期刊封面目录资料

摘要

An inductively powered implantable neural stimulator IC is presented in this paper. It features closed-loop power control, active charge balancing, and electrode impedance check functions. The stimulator IC is powered through 13.56MHz inductive link and supports 33.3kbps bi-directional telemetry with ASK for forward command transmission and LSK for backward data transmission, achieving 20V high compliance voltage, maximum 1.24mA stimulation current, and the resting potential of 50mV. The IC has an active area of 2mm×2mm implemented in 0.18-μm CMOS process with 24V LDMOS option.
机译:本文提出了一种电感可植入的可植入神经刺激器IC。它具有闭环功率控制,有源充电平衡和电极阻抗检查功能。刺激器IC通过13.56MHz电感链接和支持33.3Kbps双向遥测,并要求向前指令传输和LSK进行前向数据传输,实现20V高副规格电压,最大1.24mA刺激电流以及50mV的静止电位。 IC的有效面积为2mm×2mm,在0.18-μmCMOS过程中实现,具有24V LDMOS选项。