首页> 外文会议>IEEE Asian Solid State Circuits Conference >A 20V-compliance implantable neural stimulator IC with closed-loop power control, active charge balancing, and electrode impedance check
【24h】

A 20V-compliance implantable neural stimulator IC with closed-loop power control, active charge balancing, and electrode impedance check

机译:具有闭环功率控制,有源电荷平衡和电极阻抗检查功能的20V兼容植入式神经刺激器IC

获取原文

摘要

An inductively powered implantable neural stimulator IC is presented in this paper. It features closed-loop power control, active charge balancing, and electrode impedance check functions. The stimulator IC is powered through 13.56MHz inductive link and supports 33.3kbps bi-directional telemetry with ASK for forward command transmission and LSK for backward data transmission, achieving 20V high compliance voltage, maximum 1.24mA stimulation current, and the resting potential of 50mV. The IC has an active area of 2mm×2mm implemented in 0.18-μm CMOS process with 24V LDMOS option.
机译:本文介绍了一种感应供电的植入式神经刺激器IC。它具有闭环功率控制,有源电荷平衡和电极阻抗检查功能。刺激器IC通过13.56MHz感应链路供电,并支持33.3kbps双向遥测,其中ASK用于前向命令传输,而LSK用于后向数据传输,可实现20V的高顺从电压,最大1.24mA的激励电流和50mV的静息电位。该IC的有源面积为2mm×2mm,采用0.18μmCMOS工艺实现,并具有24V LDMOS选项。