【24h】

Pyrometer with tracking balancing

机译:高温计跟踪平衡

获取原文

摘要

Currently, one of the main metrological noncontact temperature measurement challenges is the emissivity uncertainty. This paper describes a pyrometer with emissivity effect diminishing through the use of a measuring scheme with tracking balancing in which the radiation receiver is a null-indicator. In this paper the results of the prototype pyrometer absolute error study in surfaces temperature measurement of aluminum and nickel samples are presented. There is absolute error calculated values comparison considering the emissivity table values with errors on the results of experimental measurements by the proposed method. The practical implementation of the proposed technical solution has allowed two times to reduce the error due to the emissivity uncertainty.
机译:目前,主要的计量不接触温度测量挑战之一是发射率不确定性。本文通过使用测量方案来描述具有发射率效应的高温计,其中测量方案具有跟踪平衡,其中辐射接收器是零指示器。本文提出了铝和镍样品表面温度测量的原型高温计绝对误差研究的结果。考虑到通过所提出的方法,考虑发射率表值的绝对误差计算值对比较具有关于实验测量结果结果的误差。所提出的技术解决方案的实际实施允许两次降低由于发射率不确定性而导致的误差。