Optimally Enhanced Heating for Focused Ultrasound Surgery with Split Foci, Dual-Frequency, or Multi Foci

机译:最佳地增强了聚焦超声手术的加热,分裂焦点,双频或多焦点

获取原文
获取外文期刊封面目录资料

摘要

To substantially enhance heating in HIFU treatment, several methods such as split foci, multi foci, and dual-frequency modes are used. The enhanced-cavitation heating protocols are implemented experimentally in BSA gel-phantom using four-element split-focus array. Using dual frequency of 1.2 and 2.4 MHz, the superimposing of two frequency pressures at confocal region can enhance nucleation cavitation and inertial cavitation activity. When using 135° phase shift combined with dual frequency of 1.2 and 2.4 MHz, the peak negative pressure reach maximum due to peak-negative pressures of two frequencies occur at same time, resulting strong cavitation activities. When using dual frequency of 1.2 and 2.4 MHz, 25-Hz pulse-repetition frequency (PRF), both 135° and 180° phase shift protocols, the experiment results show the largest lesion size of 10.5 x 10.5 x 11 mm~3, quickest lesion inception time of less than 0.2 s, therefore, both 135° and 180° phase shift protocols are most efficient in enhanced-cavitation heating. The filtered-PCD mean square waveforms reveal that the strong inertial-cavitation activities involve in those two treatments. The lesion size of four foci of 180° phase shift, single frequency, 25 Hz PRF, is 2 times that of 0° phase shift, single frequency even if the peak intensity of 180° case is half less than that of 0° phase shift case. When arrange multi foci using phased array in a style of a wavelength distance between neighbor foci in focal plane, the result is the same as that using split foci of 180° phase shift and single frequency.
机译:为了大大提高HIFU处理中的加热,使用若干方法,例如分裂焦点,多焦点和双频模式。增强型空化加热协议使用四元素分焦阵列实验在BSA凝胶 - 模型中实现。使用1.2和2.4MHz的双频率,共聚焦区域的两个频率压力的叠加可以增强成核空化和惯性空化活性。当使用135°相移结合双频率为1.2和2.4 MHz时,由于两个频率的峰值负压,峰值负压达到最大值同时发生,导致具有强的空化活动。当使用1.2和2.4MHz的双频率时,25赫兹脉冲重复频率(PRF),均为135°和180°相移方案,实验结果显示最大的病变尺寸为10.5 x 10.5 x 11 mm〜3,最快因此,Lesion初始时间小于0.2 s,因此,135°和180°移相方案在增强型空化加热中最有效。过滤的PCD均方波形揭示了强烈的惯性空化活动涉及这两种治疗方法。四个焦点的病变大小为180°相移,单频,25Hz prf,为0°相移的2倍,即使峰值强度为180°壳体的峰值强度为0°相移)案件。当在焦点之间的邻居焦点之间的波长距离的波长距离的样式中排列多焦点时,结果与使用180°相移和单频的分离焦点相同。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号