首页> 外文会议>Conference on Imaging, Manipulation, and Analysis of Biomolecules, Cell, and Tissues >A Comparison of Avalanche Photodiode and Photomultiplier Tube Detectors for Flow Cytometry
【24h】

A Comparison of Avalanche Photodiode and Photomultiplier Tube Detectors for Flow Cytometry

机译:雪崩光电二极管和光电倍增管检测器对流式细胞仪的比较

获取原文

摘要

Commercial flow cytometers use photomultiplier tubes (PMTs) for fluorescence detection. These detectors have high linear gain and broad dynamic range, but have limited sensitivity in the red and near infrared spectral regions. We present a comparison of avalanche photodiodes (APDs) and PMTs as detectors in flow cytometry instruments, and demonstrate improved sensitivity and resolution in the red and near infrared spectral regions using the APD. The relative performance of the PMT and APD were evaluated by simultaneously measuring the mean fluorescence intensity and coefficient of variation for emission from light emitting diode pulses, flow cytometry test beads, and fluorescently labeled cells. The relative signal to noise performance of the APD and PMT was evaluated over the 500 nm to 1050 nm wavelength range using pulsed light emitting diode light sources. While APDs have higher quantum efficiency but lower internal gain than PMTs, with appropriate external amplification the APD has signal to noise response that is comparable to PMTs in the 500 nm to 650 nm range and improved response in the 650 nm to 850 nm range The data demonstrates that the APD had performance comparable to the PMT in the spectral region between 500 to 650 nm and improved performance in the range of 650 to 1000 nm, where the PMT performance is quite poor. CD4 positive lymphocyte populations were easily identified in normal human blood both by APD and PMT using phycoerythrin labeled antibodies. In contrast, only the APD detector could resolve CD4 positive populations using 800 nm Quantum dot labeled antibodies.
机译:商业流式细胞计使用光电倍增管(PMTS)进行荧光检测。这些探测器具有高线性增益和宽动态范围,但在红色和近红外光谱区域中具有有限的灵敏度。我们向流式细胞仪器仪器中的雪崩光电二极管(APDS)和PMTS的比较,并在使用APD中展示红色和近红外光谱区域中的改善灵敏度和分辨率。通过同时测量来自发光二极管脉冲,流式细胞术试验珠和荧光标记的细胞的平均荧光强度和变化系数来评估PMT和APD的相对性能。使用脉冲发光二极管光源在500nm至1050nm波长范围内评估APD和PMT的相对信号和PMT的相对信号。虽然APDS具有更高的量子效率但内部增益较低,但是通过适当的外部放大,APD具有信号与噪声响应相当,与500nm至650nm范围的PMT相当,并且在650nm至850nm中的改善响应范围内容展示APD具有与光谱区域中的PMT相当的性能,在500至650nm之间,并且在650至1000nm的范围内提高性能,其中PMT性能相当差。使用Phycoerythrin标记的抗体,通过APD和PMT在正常人体血液中容易地鉴定CD4阳性淋巴细胞群。相比之下,只有APD检测器可以使用800nm量子点标记抗体来解决CD4阳性群体。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号