A Comparison of Avalanche Photodiode and Photomultiplier Tube Detectors for Flow Cytometry

机译:流式细胞术的雪崩光电二极管和光电倍增管检测器的比较

获取原文
获取原文并翻译 | 示例

摘要

Commercial flow cytometers use photomultiplier tubes (PMTs) for fluorescence detection. These detectors have high linear gain and broad dynamic range, but have limited sensitivity in the red and near infrared spectral regions. We present a comparison of avalanche photodiodes (APDs) and PMTs as detectors in flow cytometry instruments, and demonstrate improved sensitivity and resolution in the red and near infrared spectral regions using the APD. The relative performance of the PMT and APD were evaluated by simultaneously measuring the mean fluorescence intensity and coefficient of variation for emission from light emitting diode pulses, flow cytometry test beads, and fluorescently labeled cells. The relative signal to noise performance of the APD and PMT was evaluated over the 500 nm to 1050 nm wavelength range using pulsed light emitting diode light sources. While APDs have higher quantum efficiency but lower internal gain than PMTs, with appropriate external amplification the APD has signal to noise response that is comparable to PMTs in the 500 nm to 650 nm range and improved response in the 650 nm to 850 nm range The data demonstrates that the APD had performance comparable to the PMT in the spectral region between 500 to 650 nm and improved performance in the range of 650 to 1000 nm, where the PMT performance is quite poor. CD4 positive lymphocyte populations were easily identified in normal human blood both by APD and PMT using phycoerythrin labeled antibodies. In contrast, only the APD detector could resolve CD4 positive populations using 800 nm Quantum dot labeled antibodies.
机译:商业流式细胞仪使用光电倍增管(PMT)进行荧光检测。这些检测器具有高线性增益和宽动态范围,但在红色和近红外光谱区域的灵敏度有限。我们目前比较雪崩光电二极管(APDs)和PMTs作为流式细胞仪中的检测器,并证明使用APD在红色和近红外光谱区域中提高了灵敏度和分辨率。通过同时测量发光二极管脉冲,流式细胞术测试珠和荧光标记细胞发射的平均荧光强度和变异系数来评估PMT和APD的相对性能。使用脉冲发光二极管光源在500 nm至1050 nm波长范围内评估了APD和PMT的相对信噪性能。尽管APD比PMT具有更高的量子效率,但内部增益却较低,但通过适当的外部放大,APD的信噪比可与500 nm至650 nm范围内的PMT媲美,并且在650 nm至850 nm范围内具有更好的响应。证明了APD在500至650 nm的光谱范围内具有与PMT相当的性能,并且在650至1000 nm范围内的性能得到了改善,而PMT性能却很差。使用藻红蛋白标记的抗体通过APD和PMT均可轻松在正常人血液中鉴定出CD4阳性淋巴细胞群。相反,只有APD检测器可以使用800 nm量子点标记的抗体分辨CD4阳性种群。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号