首页> 外文会议>International Conference on Information Technology in Biomedicine >Computational Analysis of Induced Voltage on Implanted Cardiac Pacemaker's Lead by Mobile Phones
【24h】

Computational Analysis of Induced Voltage on Implanted Cardiac Pacemaker's Lead by Mobile Phones

机译:手机植入心脏起搏器铅诱导电压的计算分析

获取原文

摘要

The proposed paper deals with simulation and analysis of induced electric voltage in pacemaker lead model within the anatomical human body model due to the near-field exposure of dipole antennas covering carrier frequencies of both, GSM and UMTS mobile phone transmission technologies. The research was carried out using electromagnetic modelling based on the Finite Integration method and analysis was performed in terms of computing induced electric field distributions and induced electric voltage near to the pacemaker lead stimulation poles. The main objective was to compare levels of induced voltage on the pacemaker lead stimulation poles for both technologies, depending on the antenna distance from human torso model. The numerical results have shown that closer proximity of antenna than the distance recommended by pacemaker manufacturers could not pose higher pacemaker interference risk. Additionally, the results of estimated induced voltage have revealed that the pacemaker interference risk of recent mobile phone transmission technology is lower compared to older ones.
机译:由于偶极天线覆盖载波频率,GSM和UMTS移动电话传输技术的载波频率近场暴露,所提出的纸张涉及解剖人体体型内的起搏器引线模型中的诱导电压的仿真和分析。使用基于有限积分法进行电磁建模进行的研究,并且在计算诱导的电场分布和靠近起搏器引线刺激杆附近的诱导电压来进行分析。主要目的是根据从人躯体模型的天线距离比较两种技术的起搏器引线刺激杆上的感应电压水平。数值结果表明,天线的接近高于起搏器制造商推荐的距离不能造成更高的起搏器干扰风险。另外,估计的感应电压的结果表明,与旧的移动电话传输技术的起搏器干扰风险较低。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号