The purpose of this study is to investigate how the scattered radiation distribution in the interventional procedure room varies with changes in cranial/caudal (CRA/CAU) and right anterior oblique/left anterior oblique (RAO/LAO) gantry angulation of a C-Arm fluoroscopic system to aid in staff dose management. The primary x-ray beam of a Toshiba Infinix fluoroscopy machine was modeled using EGSnrc (DOSXYZnrc) Monte Carlo code and the scattered radiation distributions were calculated using 5 × 10~9 photons incident on the Zubal computational phantom. The Zubal phantom is derived from a CT scan of an average adult male and is anthropomorphic with internal organs. The results show that substantial changes in the scatter dose are possible for the interventionalist next to the table with Cranial/Caudal and RAO/LAO angle variations. For frontal projections the largest change with CRA/CAU angle occurs below the table height, increasing by 50% at the position of the interventionalist next to the table for a 30 degree cranial angulation compared to a caudal angulation for a beam directed toward the abdomen. The scattered radiation distribution also is shown to change with different body regions such as the chest and abdomen. A library of 3D scatter dose-rate distributions is being developed to be implemented in a scatter display system for increased staff awareness of dose levels during procedures.
机译:本研究的目的是研究介入过程室的散射辐射分布如何随着C形臂的颅/尾(CRA / CAU)和右前斜/左前斜(RAO / LAO)龙门杆的变化而变化荧光透视系统帮助员工剂量管理。使用EGSNRC(DosxyZNRC)蒙特卡罗代码建模了TOSHIBA INFINIX荧光透视机的主要X射线束,并使用入射在拟拟拟绘制模型上的5×10〜9的光子计算散射的辐射分布。绘画幽灵衍生自平均成年男性的CT扫描,并与内脏器官具有拟人。结果表明,用颅骨/尾部和RAO /老角变化,表格旁边的介入者可能发生大量变化。对于正面投影,与朝向腹部的梁的尾部颅角相比,在表格高度以下,在台高度下,在表格高度下,在表格高度下方发生了50%,在表格旁边的介入者旁边的位置。散射的辐射分布也被示出用诸如胸部和腹部的不同体积改变。正在开发一个3D散射剂量分布图书馆以在散点图系统中实现,以增加手术期间剂量水平的员工意识。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号