首页> 外文会议>Techniques and Instrumentation for Detection of Exoplanets >Phase-retrieval-based wavefront metrology for high-contrast coronography: 2. Reconstructions through a shaped pupil apodizer
【24h】

Phase-retrieval-based wavefront metrology for high-contrast coronography: 2. Reconstructions through a shaped pupil apodizer

机译:基于相位检索的波前测定的高对比度冠状动脉造影:2。通过成形瞳孔减少剂重建

获取原文

摘要

We discuss the use of parametric phase-diverse phase retrieval to characterize and optimize the transmitted wavefront of a high-contrast apodized pupil coronagraph with and without an apodizer. We apply our method to correct the transmitted wavefront of the HiCAT (High contrast imager for Complex Aperture Telescopes) coronagraphic testbed. This correction requires a series of calibration steps, which we describe. The correction improves the system wavefront from 16 nm RMS to 3.0 nm RMS for the case where a uniform circular aperture is in place. We further measure the wavefront with the apodizer in place to be 11.7 nm RMS. Improvement to the apodized pupil phase retrieval process is necessary before a correction based on this measurement can be applied.
机译:我们讨论使用参数相位数的相位检索来表征和优化具有和不具有扬声器的高对比度使瞳孔癌的透射波前。我们应用我们的方法来纠正HICAT(复杂孔径望远镜的高对比度成像)的传输波前试验用。这种校正需要​​我们描述的一系列校准步骤。对于到位均匀圆形孔的情况,校正将系统波前从16nm rms改善为3.0nm rms。我们进一步测量吸管的波前,以适当的是11.7nm rms。在基于该测量的校正之前,需要改进对沉积的瞳孔相位检索过程。