Modification of carbon paste electrode with molecularly imprinted polymer (CP-MIP) as a voltammetric sensor for creatinine has been developed. MIP was synthesized by reacting melamine, chloranil and creatinine with a mole ratio of 1:1:0.1. Creatinine was extracted from polymer chain by using hot water to form a specific imprinted for creatinine molecule. Carbon paste-MIP electrode was prepared by mixing activated carbon, solid paraffin, and MIP in a 45:40:15(w/w %) ratio. The optimum conditions of creatinine analysis by differential pulse stripping voltammetry (DPSV) using the developed electrode were the accumulation potential -1000 mV during 90 s at pH 5. The precision of the method for 0.1-0.5 μg/L creatinine was 88.7-96.3%, while the detection limit of this method was 0.0315 μg/L. The accuracy compared by spectrophotometric method was 95.3-103.6%
机译:已经开发出具有分子印迹聚合物(CP-MIP)作为用于肌酐的伏安传感器的碳糊电极的改性。通过使三聚氰胺,氯和肌酐在摩尔比为1:1:0.1的摩尔比来合成MIP。通过使用热水从聚合物链中萃取肌酐,形成用于肌酐分子的特定印迹。通过混合活性炭,固体石蜡和MIP以45:40:15(w / w%)的比例来制备碳糊状物电极。差分脉冲剥离伏安法(DPSV)使用所开发电极的最佳条件是使用显影电极的累积电位-1000mV,在pH5.0.1-0.5μg/ l肌酐的精度为88.7-96.3% ,而该方法的检测限为0.0315μg/升。分光光度法比较的准确性为95.3-103.6%