On the response of electronic personal dosimeters in constant potential and pulsed x- ray beams

机译:关于恒定电位和脉冲X射线束中电子个人剂量计的响应

获取原文

摘要

Electronic personal dosimeters (EPDs) based on solid state detectors have widely been used but some deficiencies in their response in pulsed radiation beams have been reported. Nowadays, there is not an international standard for pulsed x-ray beams for calibration or type testing of dosimeters. Irradiation conditions for testing the response of EPDs in both the constant potential and pulsed x-ray beams were established in CDTN. Three different types of EPDs were tested in different conditions in similar ISO and IEC x-ray qualities. Results stressed the need of performing additional checks before using EPDs in constant potential or pulsed x-rays.
机译:基于固态探测器的电子个人剂量计(EPDS)已被广泛使用,但已经报道了脉冲辐射束中的响应的一些缺陷。如今,没有用于校准或型剂量计测试的脉冲X射线束的国际标准。在CDTN中建立了用于测试EPD在恒定电位和脉冲X射线束中的EPD响应的辐照条件。在不同ISO和IEC X射线素质的不同条件下测试了三种不同类型的EPD。结果强调,在使用恒定电位或脉冲X射线中使用EPD之前需要进行额外检查。