首页> 外文会议>International Federation for Medical and Biological Engineering European Conference >Measurement Quality of MITS Reconstructed Tomosynthesis Slices Depending from Parameters of Images Acquiring and Reconstruction Algorithm
【24h】

Measurement Quality of MITS Reconstructed Tomosynthesis Slices Depending from Parameters of Images Acquiring and Reconstruction Algorithm

机译:根据图像获取和重建算法的参数,MITS的测量质量重建了断层切片。

获取原文

摘要

In this paper we examine how quality of MITS reconstructed slices depends on tomosynthesis acquiring total scan angle (ANG), number of input projections (N), number of reconstructed slices (NS) by measuring the modulation transfer function (MTF) of the reconstruction, and other, simulated phantoms. Our results suggest that ANG=40°, N=70, NS=150 can be the optimal combination for chest imaging, using our implementation of MITS for reconstruction.
机译:在本文中,我们研究了MITS重建切片的质量如何取决于Tomosynthesis获取总扫描角度(Ang),通过测量重建的调制传递函数(MTF),输入投影(n),重建切片(ns)的数量,和其他模拟的幽灵。我们的结果表明,Ang = 40°,n = 70,ns = 150可以是胸部成像的最佳组合,利用我们的重建的MITS实现。