首页> 外文会议>Latin American Congress on Biomedical Engineering >Design and Characterization of a Volume-Cycled Small Animal Mechanical Ventilator Coupled with a Respiratory System Model
【24h】

Design and Characterization of a Volume-Cycled Small Animal Mechanical Ventilator Coupled with a Respiratory System Model

机译:循环小型动物机械呼吸机与呼吸系统模型的设计与表征

获取原文
获取外文期刊封面目录资料

摘要

Small animals such as mice are generally used in experiments that study the behavior of the respiratory system under the influence of drugs, and accurate gas flow measurements arc necessary due to the small volumes involved. Based on this need, a mechanical ventilator for small animals was developed using a linear actuator coupled with a 3ml precision glass syringe controlled by a driver. A mechanical load representing a mouse's respiratory system consisting of a rigid plastic cylinder with internal volume of 78ml was ventilated by the machine, and its measured resistance and elastance parameters were compared to values obtained with theoretical mathematical models. A 12.7% maximum deviation was found in the model's elastance in comparison with the predicted value, and the measured resistance did not assume the expected behavior of a linear system, decreasing sharply with the increase of frequency.
机译:小鼠等小动物通常用于研究呼吸系统在药物的影响下的行为,并且由于所涉及的小容量,准确的气体流量测量是必要的。基于这种需求,使用与由驾驶员控制的3ml精密玻璃注射器联接的线性致动器开发了用于小动物的机械呼吸机。代表鼠标呼吸系统的机械负载由机器通风,由刚性塑料圆筒组成的刚性塑料圆筒,通过机器通风,并将其测量的电阻和弹性参数与具有理论数学模型获得的值进行比较。与预测值相比,在模型的弹性中发现了12.7%的最大偏差,并且测量的电阻不假设线性系统的预期行为,随着频率的增加而急剧下降。