首页> 外文会议>European Conference of the International Federation for Medical and Biological Engineering >Brain Pressure Dynamics and Control with an External Ventricular Drainage
【24h】

Brain Pressure Dynamics and Control with an External Ventricular Drainage

机译:外部心室引流脑压力动力学和对照

获取原文

摘要

In order to compensate for the disadvantages of passive drainage of cerebrospinal fluid and to get information on optimal drainage, a mechatronic extracorporal ventricular drainage system has been developed at the Helmholtz institute. A Pi-controller adjusts the position of a linear tube squeezer to obtain the target intracranial pressure value. This paper presents the developed active drainage and first measurements of the drainage on a test rig including results of an online algorithm to detect the system's hydraulic compliance.
机译:为了弥补脑脊液被动引流的缺点并获得有关最佳排水的信息,亥姆霍兹研究所开发了一种机电反应器的体外心室排水系统。 PI-Controller调节线性管挤压器的位置,以获得目标颅内压力值。本文介绍了在试验台上的发达的主动排水和第一次测量,包括在线算法的结果,以检测系统的液压顺应性。