首页> 外文会议>International Conference onanufacturing Science and Engineering >Research on Direct Calibration of Electromagnetic Flowmeter in Partially Filled Pipes
【24h】

Research on Direct Calibration of Electromagnetic Flowmeter in Partially Filled Pipes

机译:部分填充管道电磁流量计直接校准的研究

获取原文

摘要

The methodology of the direct calibration for the electromagnetic flowmeter in partially filled pipes is presented in this paper. Based on the principle of the electromagnetic flowmeter, two main respects, including the flow liquid level and the weight function in partially filled pipes are introduced briefly, and the calibration technologies of the sensor, which consist of the device arrangements, the accurate measurement of the liquid level and the calibration of the flow discharge, are emphasized on, with the methodology based on the piecewise interpolation employed. The experimental results demonstrate that the methodology of the direct calibration is simple and effective for the real flow calibration of the electromagnetic flowmeter in partially filled pipes, with the stable measurement and the accuracy requirement satisfied.
机译:本文提出了部分填充管中的电磁流量计的直接校准的方法。基于电磁流量计的原理,简要介绍了两个主要方面,包括流动液位和在部分填充的管道中的重量函数,以及传感器的校准技术,包括设备布置,准确测量液位和流量放电的校准,基于采用分段插值的方法。实验结果表明,直接校准的方法是简单且有效的电磁流量计在部分填充管中的实际流动校准,具有稳定的测量和满足精度要求。