首页> 外文会议>International Conference on Biomedical Electronics and Devices >AN ELECTRONIC INTERFACE FOR NEURAL ACTIVITY RECORDING AND STIMULATION
【24h】

AN ELECTRONIC INTERFACE FOR NEURAL ACTIVITY RECORDING AND STIMULATION

机译:用于神经活动记录和刺激的电子界面

获取原文

摘要

A portable neural activity acquisition and stimulation system by means of tfLIFE implantable electrodes has been realized. The detecting circuit provides: a selective filtering made up of a 4th order high pass Multiple Feedback filter (f_(-3dB) = 1.1kHz) and a 4th order low pass Multiple Feedback filter (f_(-3dB) = 2kHz), a variable gain (24dB - 44dB) and a 16 bit analog to digital conversion. The stimulator allows to generate specific electrical signals through a digital-to-analog converter while stimulation parameters as frequency, duration and intensity are controlled by a digital microcontroller. Simulation results and first experimental results of the interface demonstrate how neural signals of a few of microvolts can be filtered, programmable amplified and digitalized without distortion.
机译:通过TFLIFE可植入电极实现了便携式神经活动采集和刺激系统。检测电路提供:由第4阶PORV PART多次反馈滤波器组成的选择性滤波(F _( - 3DB)= 1.1kHz)和第4阶低通过多反馈滤波器(f _( - 3db)= 2khz),一个变量增益(24dB - 44dB)和16位模数转换。刺激器允许通过数字到模拟转换器产生特定的电信号,同时刺激参数作为频率,持续时间和强度由数字微控制器控制。仿真结果和界面的第一个实验结果表明,可以过滤少数微伏的神经信号,可编程放大和数字化而没有失真。