Impact of radiation dose level on CT number accuracy in photon counting CT

机译:辐射剂量水平对光子计数CT中CT数精度的影响

获取原文

摘要

It is well known that when radiation dose is reduced in x-ray computed tomography, image noise increases and noise induced streaks may appear. However, to quantify the accuracy of the image reconstruction, another source of error, known as the statistical bias, also needs to be considered. In the projection domain, signal bias originates in the quantum nature of the x-ray photon fluctuations as well as the nonlinear nature of the logarithmic transform used to generate line integral data. The bias in the projection domain is then propagated into the image domain through the image reconstruction process to generate CT number biases. The purpose of this work is to experimentally study the dependence of CT number bias on the radiation dose.
机译:众所周知,当辐射剂量在X射线计算机断层扫描中减小时,可以出现图像噪声增加和噪声感应条纹。然而,为了量化图像重建的准确性,还需要考虑另一个称为统计偏差的误差源。在投影域中,信号偏置源自X射线光子波动的量子性质以及用于生成线积分数据的对数变换的非线性性质。然后通过图像重建处理将投影域中的偏置传播到图像域中以产生CT数偏差。这项工作的目的是通过实验研究CT号偏差对辐射剂量的依赖性。