Impact of radiation dose level on CT number accuracy in photon counting CT

机译:光子计数CT中辐射剂量水平对CT数准确性的影响

获取原文

摘要

It is well known that when radiation dose is reduced in x-ray computed tomography, image noise increases and noise induced streaks may appear. However, to quantify the accuracy of the image reconstruction, another source of error, known as the statistical bias, also needs to be considered. In the projection domain, signal bias originates in the quantum nature of the x-ray photon fluctuations as well as the nonlinear nature of the logarithmic transform used to generate line integral data. The bias in the projection domain is then propagated into the image domain through the image reconstruction process to generate CT number biases. The purpose of this work is to experimentally study the dependence of CT number bias on the radiation dose.
机译:众所周知,当在X射线计算机断层摄影术中减少辐射剂量时,图像噪声增加并且可能出现噪声引起的条纹。但是,为了量化图像重建的准确性,还需要考虑另一个误差源,称为统计偏差。在投影域中,信号偏差源自x射线光子涨落的量子性质以及用于生成线积分数据的对数变换的非线性性质。然后,通过图像重建过程将投影域中的偏差传播到图像域中,以生成CT数偏差。这项工作的目的是通过实验研究CT数偏差对辐射剂量的依赖性。