Simulation of Temperature Field Induced by 8-Element Phased Array HIFU Transducer with Concave Spherical Surface

机译:8元分阵阵列HIFU传感器用凹形球面诱导温度场的仿真

获取原文

摘要

Multi-element High Intensity Focused Ultrasound (HIFU) transducers can change their focal lengths and form multi-foci. In this paper the Westervelt formula and Pennes bio-heat transfer equation have been used along, with the Finite Difference Time Domain (FDTD) method, to study the temperature distribution induced by an 8-element phased array HIFU transducer inside the human body. We evaluated the effects of the gap in the arc between two rings, the frequency of excitation function and pre-focal length on the temperature field. For HIFU therapy, skin burns were caused by high frequency, small pre-focal length, or a big gap between two rings. The focal region may be no longer an ellipsoid due to high frequency. In addition, the actual focal length is slightly different from the pre-focal length.
机译:多元素高强度聚焦超声(HIFU)换能器可以改变其焦距并形成多焦点。在本文中,威斯特察雷特公式和彭恩斯生物传热方程已经使用,其中有限差分时域(FDTD)方法研究了通过人体内的8元件相位阵列HIFU换能器引起的温度分布。我们评估了两个环之间的弧形中间隙,激发功能频率和温度场的焦距。对于HIFU治疗,皮肤燃烧是由高频,小焦距的焦距或两个环之间的差距引起的。由于高频率,焦点区域可能不再是椭圆形。另外,实际焦距与前焦距略有不同。