Biomechanical Characterization of a Cervical Corporectomy using Porcine Specimens, following an Experimental Approach

机译:一种实验方法,使用猪标本的宫颈共计术的生物力学表征

获取原文

摘要

In this paper, the interaction among cervical vertebrae, a cervical plate and a bone graft implant, which is developed in a Corporectomy, is analyzed in an experimental form. In the case of specific damaged vertebra, its replacement is one of the alternative solutions. However, the displacement between the vertebral adjacent facets and the bone graft is a critical parameter which has to be evaluated in order to ensure the stability of the spine. Besides, it is advisable to make a precise evaluation of the structural integrity of the arrangement. For this study, porcine cervical vertebrae (C3-C5) were instrumented in order to replace a damaged C4 vertebra. This arrangement was tested under compression. The experimental observations were complemented with a numerical model. The displacements between the vertebral facets and the bone graft were measured. They are lower than 3 mm in order to develop stability in the spine. Besides, the proposed arrangement has structural integrity and the surgical procedure is simplified, as no wires are used.
机译:本文以实验形式分析了宫颈椎骨,宫颈板和骨移植植入物之间的相互作用。在特定损伤的椎骨的情况下,其更换是替代解决方案之一。然而,椎体相邻刻面和骨移植物之间的位移是必须评估的关键参数,以确保脊柱的稳定性。此外,建议准确评估安排的结构完整性。对于该研究,猪颈椎(C3-C5)被仪器化,以取代受损的C4椎骨。这种布置在压缩下进行了测试。实验观察与数值模型互补。测量椎体套头和骨移植物之间的位移。它们低于3毫米,以便在脊柱中产生稳定性。此外,所提出的布置具有结构完整性,并且简化了外科手术,因为没有使用电线。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号