Quantification of Radiographic Image Quality Based on Patient AnatomicalContrast-to-Noise Ratio - A Preliminary Study with Chest Images

机译:基于患者的患者解剖学区集对噪声比的放射线图像质量的量化 - 胸部图像初步研究

获取原文

摘要

The quality of a digital radiograph for diagnostic imaging depends on many factors, such as the capture system DQE and MTF, the exposure technique factors, the patient anatomy, and the particular image processing method and processing parameters used. Therefore, the overall image quality as perceived by the radiologists depends on many factors. This work explores objective image quality metrics directly from display-ready patient images. A preliminary study was conducted based on a multi-frequency analysis of anatomy contrast and noise magnitude from 250 computed radiography (CR) chest radiographs (150 PA, 50 AP captured with anti-scatter grids, and 50 AP without grids). The contrast and noise values were evaluated in different sub-bands separately according to their frequency properties. Contrast-Noise ratio (CNR) was calculated, the results correlated well with the human observers' overall impression on the images captured with and without grids.
机译:用于诊断成像的数字射线照片的质量取决于许多因素,例如捕获系统DQE和MTF,曝光技术因素,患者解剖学和特定图像处理方法和使用的处理参数。因此,放射科医师感知的整体图像质量取决于许多因素。这项工作直接从显示就绪患者图像探讨了客观图像质量指标。基于从250计算的射线照相(CR)胸部射线照片的解剖学对比度和噪声幅度的多频分析进行了初步研究(150Pa,用反散射网格捕获的50个AP,以及50个没有网格的AP)。根据其频率属性单独在不同的子带中评估对比度和噪声值。计算对比度噪声比(CNR),结果与人类观察者的整体印象良好,在没有网格上捕获的图像上的图像的整体印象。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号