首页> 外文会议>Test and Measurement Conference >Wavelength References at the NMISA Fibre Optic Laboratory to Improve Uncertainty of Measurement for Telecommunication Wavelengths
【24h】

Wavelength References at the NMISA Fibre Optic Laboratory to Improve Uncertainty of Measurement for Telecommunication Wavelengths

机译:NMISA光纤实验室的波长参考,以改善电信波长测量的不确定性

获取原文

摘要

Absolute wavelength references, hydrogen cyanide (HCN) and carbon monoxide (CO) in a glass gas cell coupled with fibre optic ports is used to get reliable and more accurate wavelength reference to improve uncertainties for fibre optic wavelength calibrations in the S-, C, and L-bands for telecommunication purposes. Method validation using a gas cell and a calibrated laser are discussed. The accuracy of the Optical Spectrum Analyser (OSA) is determined and wavelength sources can now be calibrated with an uncertainty of 0,05 nm at NMISA.