首页> 外文会议>Clinical Histocompatibility Workshop. >POSITIVE FLOW CYTOMETRY CROSSHATCH DUE TO LOW LEVELS OF Bw6 ANTIBODIES DETECTED BY LABScreen~R ASSAY
【24h】

POSITIVE FLOW CYTOMETRY CROSSHATCH DUE TO LOW LEVELS OF Bw6 ANTIBODIES DETECTED BY LABScreen~R ASSAY

机译:由于Labscreen〜R测定检测到的低水平BW6抗体,阳性流式细胞术交叉液

获取原文

摘要

This case report focuses on a sensitized kidney transplant candidate with weak reactivity against several HLA-B locus antibodies as determined with LABScreen~R. Each reactive bead expressed the HLA-Bw6 epitope. Although the detected antibodies are below our positive cut-off threshold, it led to a strong positive crossmatch. This case report suggests that the LABScreen~R assay can detect HLA antibodies generally considered too weak to result in a positive crossmatch and careful consideration should be given to the pattern of reactivity of beads. Signal strength alone is not a sufficient metric to assess the potential of a so-called weak antibody to result in a positive corssmatch.
机译:本病例报告侧重于敏感的肾移植候选候选较弱的反应性,与鼠屏〜r测定的几种HLA-B座位抗体。每个反应性珠表达HLA-BW6表位。虽然检测到的抗体低于我们的正截止阈值,但它导致了强烈的正交交叉迁移。本病例报告表明,Labscreen〜R测定可以检测通常认为过于弱的HLA抗体,以导致阳性交叉匹配,并且应仔细考虑珠子的反应性模式。单独的信号强度不是足够的指标,以评估所谓的弱抗体的电位,以导致阳性辅助斯坎斯。