The Immersion Profile in a Triple Point of Water Fixed Point Cell for Different SPRTs of the Same Type

机译:用于不同运动的不同运动的水固定点电池三重点中的浸入轮廓

获取原文

摘要

The defining fixed-points of the International Temperature Scale of 1990 (ITS-90) are reference points for temperature calibration. The measured temperature inside the fixed-point cell depends on thermometer immersion, since measurements are made below the surface of the fixed-point material and additional effect of the hydrostatic pressure has to be taken into account. Also, the heat flux along the thermometer stem can affect the measured temperature. Since the measurement of the immersion profile is rather lengthy and time consuming (several hours for measurements at single temperature), the laboratories measure it for every type of SPRT calibrated, even only for particular SPRT design, but not for every thermometer under calibration. Even the small differences in manufacturing procedure can contribute to the different response of the particular SPRT of the same type to the immersion profile. Also, usage of the thermometer, measurement set up, can change its characteristics and response to the immersion profile. Thus the uncertainty budget of the thermometer due to perturbing heat exchanges is at the end different to every particular thermometer and measurement set-up, and should be determined.
机译:1990年(ITS-90)的国际温度尺度的定义固定点是温度校准的参考点。固定点电池内的测量温度取决于温度计浸没,因为在固定点材料的表面下测量,并且必须考虑静液压的附加效果。而且,沿着温度计杆的热通量可以影响测量的温度。由于浸入曲线的测量相当长,并且在单温时测量几个小时),因此实验室为每种类型的SPRIBRET测量,即使仅用于特定的SPRT设计,但不适用于每个校准的温度计。即使制造过程的小差异也可以有助于与浸入曲线相同类型的特定SPRT的不同响应。此外,温度计使用测量设定的使用可以改变其特征和对浸入曲线的响应。因此,由于扰动热交换器导致的温度计的不确定性预算是与每个特定温度计和测量设置不同的末端,并且应该确定。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号