PERFORMANCE EVALUATION OF AN AUTOMATED THREE-STAGE CRYOGENIC PRECONCENTRATOR MODULE FOR THE COMPOUND-SPECIFIC CARBON ISOTOPE ANALYSIS OF BENZENE, ISOPRENE , N-PENTANE, METHACROLEIN, METHYL VINYL CHLORIDE AND TOLUENE

机译:苯,异戊二烯,正戊烷,甲基丙烯醛,甲基氯乙烯和甲苯的复合特异性三阶段低温预浓缩剂模块的性能评价

获取原文

摘要

The instrument specification on precision reported by manufacturers of gas chromatograph-combustion isotope ratio mass spectrometers (GC-C-IRMS) is generally 0.5%o. By comparing the conventional spit and splitless GC injection techniques to on-column injection a decrease in the standard deviation better than 0.3%o for the majority of analytes was reported. In this paper, we present precision values with a standard deviation as little as 0.05‰ measured within the recommended linear range of 1 to 10 nA of the IRMS by using the cooled injector system (CIS 4) (GERSTEL GmbH & Co. KG, Germany) operated in the solvent vent mode. The sample trapping and thermal desorption was achieved using the Thermo Desorption-Large system (TDSG-Large -GERSTEL GmbH & Co. KG, Germany) functioning as part of a fully automated cryogenic preconcentrator/focuser module (CP/F) geared at preconcentrating air volumes as little as 0.1 L and as large as 178 L. The compounds that were tested were benzene, isoprene, methacrolein (MACR), methyl vinyl ketone (MVK), n-pentane and toluene. In addition, the difference between on-line and off-line carbon isotope composition values was determined for all six compounds. The purpose of this work was to test the potential of the new module in terms of reproducibility and accuracy regarding isotope composition for sample volumes up to 178 L after optimizing water trapping temperatures, sample trapping and focusing temperatures, peak resolution between n-pentane and isoprene and the dynamic and static set points and times needed in the sample preparation process using the Gerstel TDSG Large Software (Version 0.5.3.5).
机译:气相色谱燃烧同位素比例质谱仪(GC-C-IRMS)制造商报道的精度仪表规范通常为0.5%O。通过将常规吐痰和无分裂的GC喷射技术与在柱注射中进行比较,报道了大多数分析物的标准偏差的降低。在本文中,我们通过使用冷却的喷射器系统(CIS 4)(Gersel GmbH&Co.KG,德国Gersel GmbH&Co.KG )在溶剂通气模式下操作。使用热解吸 - 大型系统(TDSG-MATRY-GMBH&Co.KG,Germany)作为全自动低温预融合器/聚焦器模块(CP / F)的一部分,实现了样品捕获和热解吸,以预先介入空气高达0.1L的体积和大约178L。测试的化合物是苯,异戊二烯,甲基丙烯(MacR),甲基乙烯基酮(MVK),正戊烷和甲苯。此外,确定所有六种化合物确定在线和离线碳同位素组成值之间的差异。本作品的目的是在优化水捕获温度,样品捕获和聚焦温度,正戊烷和异戊二烯之间的样品体积的再现性和准确性方面的再现性和准确性来测试新模块的潜力。使用Gerstel TDSG大型软件(版本0.5.3.5)样品准备过程中的动态和静态设定点和时间。

著录项

相似文献

  • 外文文献
  • 中文文献
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号