Investigation of a Hydrodynamic Performance of a Ventricular Assist Device After Its Long-Term Use in Clinical Application

机译:在临床应用中长期应用后心室辅助装置的流体动力学性能研究

获取原文

摘要

A long-term durability of a ventricular assist device (VAD) is required due to a shortage of donor hearts for cardiac transplantation, but there is no analyzed pump data after long-term use. This study aimed to perform a comparative study between new VAD and VADs after long-term use. The hydrodynamic performance of the used Toyobo VADs (mean period of 5 months with the maximum of 12 months) was evaluated in a mock circulatory system, where a new VAD was used as a control. Although a remarkable difference was not observed in terms of mean flow rate, flow and pressure waveforms varied significantly. Then, the pressure-volume relationship of each pump was measured. It was found that the capacity of long-term VADs was reduced. Although a further study is required, these results suggested that a long-term use of VAD may cause a change in mechanical properties of polymer materials.
机译:由于心脏移植的供体心脏短缺,需要长期耐久性(VAD),但在长期使用后没有分析的泵数据。本研究旨在在长期使用后对新VAD和VAD进行比较研究。在模拟循环系统中评估了二手Toyobo VAD的流体动力学性能(平均5个月,最多12个月),其中新VAD用作对照。虽然在平均流速方面未观察到显着差异,但流量和压力波形显着变化。然后,测量每个泵的压力量关系。发现长期VAD的能力降低。尽管需要进一步的研究,但这些结果表明,长期使用VAD可能导致聚合物材料的力学性能变化。