Nitinol self-expanding stents are used to treat peripheral occluded vessels such as the superficial femoral artery or the carotid. The complex vessel articulation requires a stent device that is flexible and kink resistant yet durable. The present study shows how the latest advances in commercially available engineering software tools permit engineering simulations of the many aspects of the Nitinol stent design and analysis. Two stent geometries are evaluated: a helical type stent design, and a more traditional straight strut, with multiple crowns design. The fatigue performance of the two stents is compared. The results show that advanced nonlinear finite element simulations and fatigue predictions of the Nitinol stent are possible today inside realistic simulated human arteries. The finite element analysis software used in this study is SimXpert, Marc, and Mentat (MSC Software, Santa Ana, CA).
机译:Nitinol自膨胀支架用于治疗外围闭塞容器,例如表面股动脉或颈动脉。复杂的血管铰接需要一块支架装置,该支架装置是柔性的,扭结但耐用的。本研究表明,商业上可获得的工程软件工具的最新进展情况允许工程模拟镍钛合金支架设计和分析的许多方面。评估两个支架几何形状:螺旋式支架设计,以及更传统的直支撑,具有多个冠设计。比较了两个支架的疲劳性能。结果表明,今天在现实的模拟人类动脉内,奈多尔支架的先进非线性有限元模拟和抗脂肪预测是可能的。本研究中使用的有限元分析软件是Simxpert,Marc和Mentat(MSC软件,Santa Ana,CA)。