首页> 外文会议>Cryogenic Engineering Conference >SNS Cryogenic Test Facility Kinney Vacuum Pump Commissioning and Operation at 2 K
【24h】

SNS Cryogenic Test Facility Kinney Vacuum Pump Commissioning and Operation at 2 K

机译:SNS低温测试设施Kinney真空泵调试和操作在2 k

获取原文

摘要

The Spallation Neutron Source (SNS) at Oak Ridge National Laboratory (ORNL) has built and commissioned an independent Cryogenic Test Facility (CTF) in support of testing in the Radio-frequency Test Facility (RFTF). Superconducting Radio-frequency Cavity (SRF) testing was initially conducted with the CTF cold box at 4.5 K. A Kinney vacuum pump skid consisting of a roots blower with a liquid ring backing pump was recently added to the CTF system to provide testing capabilities at 2 K. System design, pump refurbishment and installation of the Kinney pump will be presented. During the commissioning and initial testing period with the Kinney pump, several barriers to achieve reliable operation were experienced. Details of these lessons learned and improvements to skid operations will be presented. Pump capacity data will also be presented.
机译:橡树岭国家实验室(ORNL)的介绍中子源(SNS)已经建立和委托了一个独立的低温测试设施(CTF),以支持在射频测试设施(RFTF)中进行测试。 最初用4.5 K的CTF冷箱进行超导射频腔 K.系统设计,泵翻新和Kinney Pump的安装将显示。 在调试和初始测试期间与Kinney Pump,经历了几种达到可靠运行的障碍。 将提出这些经验教训的详细信息和对滑雪操作的改进。 泵容量数据也将出现。