摘要

This study was performed for a purpose of developing a passive sampler, which can quantify formaldehyde in indoor air by sampling for about 1 hour. This study was investigated by using the method described as follows: Formaldehyde reacts with 1-hydrazinophthazine to produce s-triazolo [3,4-a] -phthalazine, formaldehyde concentration can be measured from fluorescence of the production by High Performance Liquid Chromatogaph with fluorescence detector. Passive sampler packed with alkali-impregnated silica gel of 0.8g in a porous tube made from polytetrafluoroethylene resin was prepared. The uptake rate of the passive sampler is 3.06 mu. g/ppm/hr, which is equivalent to 42ml/min of sampling rate of a pump method. The repeatablity of the passive sampler is 6.9 percent in coefficient of variation. The passive sampler can be measured up to 10 ppb of formaldehyde with 1 hour sampling. The measured value of present method and 2,4-dinitrophenylhydrazine active sampler method agreed well.
机译:该研究是为了开发被动采样器的目的,通过取样约1小时来量化室内空气中的甲醛。通过使用如下所述的方法研究了该研究:甲醛与1-肼噻嗪反应以产生S-三唑[3,4-A] - 苯噻嗪,可以通过高性能液体染色物与荧光产生的荧光从生产的荧光测量甲醛浓度探测器。制备用碱浸渍硅胶填充的被动采样器在由聚四氟乙烯树脂制成的多孔管中0.8g。被动采样器的摄取率为3.06亩。 g / ppm / hr,它相当于泵方法的42ml / min。无源采样器的重复性在变异系数中为6.9%。无源取样器可以测量高达10ppb的甲醛,用1小时取样。本方法的测量值和2,4-二硝基苯基肼活性采样器方法吻合良好。