The BL07A beamline in New SUBARU, University of Hyogo, has been used for many studies of new materials. A new double crystal monochromator controlled by integrated computing was designed and installed in the beamline in 2014. In this report we will discuss the unique features of this new monochromator, MKZ-7NS. This monochromator was not designed exclusively for use in BL07A; on the contrary, it was designed to be installed at low cost in various beamlines to facilitate the industrial applications of medium-scale synchrotron radiation facilities. Thus, the design of the monochromator utilized common packages that can satisfy the wide variety of specifications required at different synchrotron radiation facilities. This monochromator can be easily optimized for any beamline due to the fact that a few control parameters can be suitably customized. The beam offset can be fixed precisely even if one of the two slave axes is omitted. This design reduces the convolution of mechanical errors. Moreover, the monochromator's control mechanism is very compact, making it possible to reduce the size of the vacuum chamber can be made smaller.
机译:兵库大大学新斯巴鲁的BL07A梁线已被用于许多新材料的研究。通过集成计算控制的新的双晶单色器是在2014年的梁线上设计的。在本报告中,我们将讨论这款新单色器MKZ-7ns的独特功能。该单色器不是专门用于BL07A的设计;相反,它旨在以低成本安装在各种光束线上,以便于中型同步辐射设施的工业应用。因此,单色仪的设计利用了可满足不同同步辐射设施所需的各种规格的公共包装。由于可以适当地定制一些控制参数,因此可以为任何光束线容易地优化该单色器。即使省略了两个从轴中的一个,也可以确定地固定光束偏移。这种设计减少了机械误差的卷积。此外,单色仪的控制机构非常紧凑,使得可以减小真空室的尺寸可以使得可以更小。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号