In the particular study, the micropatterning of the Ti5A14V alloy in as-received, solution treated and solution and precipitated condition was done using electron beam surface modification (EBSM) by scanning electron beam. The nanostructured TiN/TiC>2 films were deposited over the EBSM Ti5A14V substrates using dc magnetron sputtering and cathodic arc and glow-discharge methods. The electrochemical potentiodynamic tests were performed in Ringer saline at 37±0.05 °C and a pH of 5.7 in order to determine the corrosion resistance behavior of the coated and uncoated EBSM substrates as compared with the bare as-received, solution treated and solution and precipitated samples. The presented results gave an idea of the corrosion behavior of the examined implant systems in aggressive environment and external conditions close to the body right.
机译:在特定的研究中,通过扫描电子束使用电子束表面改性(EBSM)完成接收的溶液处理和溶液和沉淀条件的Ti5a14V合金的微图案。使用DC磁控溅射和阴极弧和辉光排出方法在EBSM Ti5a14V基板上沉积纳米结构罐/ TiC> 2膜。在37±0.05℃和pH为5.7的pH值中,在林果盐水中进行电化学电位试验,以确定涂覆和未涂覆的EBSM底物的耐腐蚀性,与裸露的接收溶液处理和溶液相比并沉淀样品。所提出的结果介绍了所检查的植入系统在侵略性环境中的腐蚀行为和靠近身体的外部条件。