Parametric Design Study of a Long Axial Field-of-View PET Scanner using a Block-Detector Tomograph Simulation of a Cylindrical Phantom

机译:使用圆柱形模拟的块探测器断层扫描仪模拟长轴视野PET扫描仪的参数化设计研究

获取原文

摘要

We are conducting a two-phase study, which aims to select design parameters of a long-axis positron emission tomography (PET) scanner that are a compromise between detection performance and cost. In this first phase, we examine the effects of axial length, detector thickness and collimator geometry on the noise equivalent count rate per axial length (λNEC) and noise equivalent count rate per slice (dNEC). We use these metrics as approximate, but quickly computed, indicators of a PET-scanner's performance at a detection task. From this first phase, we select a subset of scanner designs for which we can conduct a detailed study of tumor detectability and quantitation accuracy in whole-body PET imaging.
机译:我们正在进行两阶段的研究,旨在选择长轴正电子发射断层扫描(PET)扫描仪的设计参数,这些参数是检测性能和成本之间的折衷。在该第一阶段,我们检查轴向长度,探测器厚度和准直器几何形状对每个轴向长度(λnec)和噪声等效计率(DNEC)的噪声等效计率的影响。我们将这些指标用作近似,但快速计算,宠物扫描仪在检测任务中的性能的指标。从这个第一阶段,我们选择扫描仪设计的子集,我们可以在全身宠物成像中进行详细研究肿瘤可检测性和定量精度。