The role of breast compression in scintimammography

机译:乳房压缩在ScintImampicoct中的作用

获取原文

摘要

Phantom model was performed to study the effect of breast compression on signal-to-noise ratio (SNR) for a dedicated high-resolution gamma camera (Single Photon Emission Mammography, or 'SPEM') and a conventional one as typically employed in prone scintimammography. The phantom was designed to simulate the effects of lesion size and of scatter from nearby torso activity. The phantom studies showed that lesions SNR was higher with the SPEM camera than with the conventional camera, and that SNR was always improved with compression for both cameras. Since the stage of breast cancer diagnosis affects patient prognosis, it is important to optimize breast examinations for small (i.e., T1a and T1b) lesions. For one-cm size lesions (clinical stage Tic), SNR was maximized when compression was less than 12 cm, and little additional benefit was derived from further compression. For subcentimeter (clinical stage T1b) lesions, SNR was maximized when compression was less than 6 cm. These data are consistent with a short clinical study in which detection sensitivity for small cancers was improved with the SPEM camera as compared to a conventional gamma camera. We conclude that, in order to image early breast cancers (stage T1b), it is important to apply breast compression.
机译:进行幻影模型以研究乳房压缩对专用高分辨率伽马摄像机(单光子发射乳房X线摄影摄像机(单光子发射乳房X线摄像觉或'SPEM')的信噪比(SNR)的效果,以及通常在易于瞬息术中使用的传统型。幻影旨在模拟病变大小的影响和从附近的躯干活动的散射。幻影研究表明,与传统相机的贴片相机损伤SNR较高,并且SNR随着两个摄像机的压缩而始终得到改善。由于乳腺癌诊断的阶段影响患者预后,重要的是优化小(即T1A和T1B)病变的乳房检查。对于单厘米尺寸的病变(临床阶段TIC),当压缩小于12cm时,SNR最大化,并且从进一步的压缩得出几乎没有额外的益处。对于子心尖计(临床阶段T1B)病变,当压缩小于6cm时,SNR最大化。这些数据与短暂的临床研究一致,其中与传统的伽马摄像机相比,使用SPEM相机改善了小癌症的检测灵敏度。我们得出结论,为了图像早期乳腺癌(阶段T1B),应用乳房压缩是重要的。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号