首页> 外文会议>Conference on Percision Electromagnetic Measurements >Stability and accuracy results with the on/METAS continuous cesium fountain
【24h】

Stability and accuracy results with the on/METAS continuous cesium fountain

机译:稳定性和准确性结果与on / Metas连续铯喷泉

获取原文

摘要

The construction of a primary frequency standard based on a continuous beam of laser cooled cesium atoms has been completed. The device is operational and undergoes its first accuracy evaluation. The paper presents stability measurements and accuracy issues, with a special emphasis on the uncertainty contributions that differ from a pulsed fountain.
机译:已经完成了基于连续的激光冷却铯原子的初级频率标准的构造。该装置是运行的,并经历其第一精度评估。本文提出了稳定性测量和准确性问题,特别强调了与脉冲喷泉不同的不确定性贡献。