CALIBRATION OF THE MOESSBAUER SPECTROMETER VELOCITY BY OPTICAL METHODS

机译:光学方法校准Moessbauer光谱仪速度

获取原文

摘要

There is basically no limit at the high velocities as far as the Mossbauer spec-troscopy is considered. Very low velocities require low repetition frequencies of the spectrometer and their calibration suffers from the low count rate of the TTL pulses. Due to the fact that a basic laser frequency cannot be increased significantly one has to use multiple scattering with a well defined index k. However, standard multiple scattering interferometers, e.g., Fabry-Perot or Lummer-Gehrcke devices cannot be used as the multiplicity k is poorly defined for them.
机译:只要考虑Mossbauer Spec-Troscopy,高速度基本上没有限制。非常低的速度需要光谱仪的低重复频率,并且它们的校准患有TTL脉冲的低计数率。由于基本激光频率不能显着增加一个,必须使用具有良好定义的索引k的多个散射。然而,标准多个散射干涉仪,例如法布里 - 珀罗或Lummer-Gehrcke装置不能用作多重k为它们定义很差。