A prosthetic disc nucleus device has been designed to replace the degenerated lumbar intervertebral disc (IVD) nucleus. In this study, we measured the creep response of a lumbar motion segment in the intact condition, after a nucleotomy, and after implantation with the device. The creep behavior was modeled with a three parameter standard linear solid to determine the elastic (E_1, E_2) and viscous (#eta#) constants that describe the behavior of the disc. The average elastic damping constant (E_1), and the viscous constant (#eta#) were significantly lower after the nucleotomy than in the intact state. There were no significant differences between the intact motion segment and the device-implanted motion segment. This study demonstrates that the prosthetic disc nucleus device restores the viscoelastic behavior of the intact spine after nucleotomy.
机译:假体圆盘核装置已经设计成替代退化的腰椎椎间盘(IVD)核。在这项研究中,我们在核细胞之后和植入装置的植入后测量了腰椎运动段的蠕变响应。蠕变行为用三个参数标准线性固体进行建模,以确定描述光盘行为的弹性(E_1,E_2)和粘性(#ETA#)常量。在核桃术后,平均弹性阻尼常数(E_1)和粘性常数(#ETA#)显着低于完整状态。完整的运动段和器件植入的运动段之间没有显着差异。本研究表明,假体椎间盘核电片核心核核电片恢复了核桃术后完整脊柱的粘弹性行为。