首页> 外文会议>IEEE International Conference Image Processing >Timing specifications and accuracy of the real-time 3D echocardiographic reconstruction system
【24h】

Timing specifications and accuracy of the real-time 3D echocardiographic reconstruction system

机译:实时3D超声心动图重建系统的时间规范和准确性

获取原文

摘要

We analyze the performance of the real-time system for 3D reconstruction of cardiac structures from successive 2D B-scan ultrasound images acquired using the Tilt Echo technique, developed by Buckey et al. (1987, 1993). This system is used to evaluate cardiac performance parameters such as stroke volume and ventricular mass. The design of such a real-time system involves consideration of two features, viz. the accuracy of the 3D reconstructions generated by the system, and the timing behavior of the system. We describe the procedure to test the accuracy of the system and present results to demonstrate the accuracy of reconstruction. We analyze the timing behavior of the system based on simulations using Modechart. We verify that the system design satisfies the real-time constraints.
机译:我们分析了使用倾斜回波技术获取的连续的2D B扫描超声图像的3D重构实时系统的性能,由Buckey等人开发。 (1987,1993)。该系统用于评估心脏性能参数,例如行程体积和心室质量。这种实时系统的设计涉及考虑两个特征,viz。系统生成的3D重建的准确性,以及系统的定时行为。我们描述了测试系统准确性的程序和目前的结果,以证明重建的准确性。我们根据使用ModeChart的模拟分析系统的时序行为。我们验证系统设计是否满足了实时约束。