We investigated the effect of varying fluence and pulse width on the ablation rate and consequent thermal damage of the Ho:YAG (2.130 $mu@m) laser. The rate of ablation on fresh bovine knee joint tissues, fibrous cartilage, hyaline cartilage, and bone in saline was determined after varying the fluence (160 - 640 J/cm$+2$/) and pulse width (150, 250, 450 $mu@sec, FWHM) at a repetition rate of 2 Hz. A 400/440 $mu@m fiber was used. The ablation rate increased linearly with the fluence. In fibrocartilage, different pulse durations generated significant changes in the ablation rates, but showed minor effects on hyaline cartilage and bone. The heat of ablation for all three tissue types decreased after lengthening the pulse.
机译:我们调查了不同流量和脉冲宽度对烧蚀速率的影响,并因此的HO:YAG(2.130 $ MU @ M)激光器的热损坏。在不同流量的情况下测定新鲜牛膝关节组织,纤维软骨,透明软骨和盐水中的骨质纤维软骨和骨率(160-640J / cm $ + 2 $ /)和脉冲宽度(150,250,450 $ MU @ SEC,FWHM)以2 Hz的重复率。使用400/440 $ MU @ M光纤。消融率随着注速而线性增加。在纤维纤维下,不同的脉冲持续时间产生了烧蚀率的显着变化,但对透明软骨和骨骼显示出微小的影响。在延长脉冲后,所有三种组织类型的消融热量降低。